You can run the Cellazyme method at pH 6.2. We suggest that you use a Maleate buffer (prepared from maleic acid at 25 mM concentration). We would also suggest that you stop the reaction with tri-sodium phosphate (adjusted to pH 11.0). This will however affect the standard curve as different parameters are being used. A new standard curve can be generated using your own conditions by standardising the activity of a purified endo-cellulase on say CM-Cellulose 4M with Nelson/Somogyi reducing sugar method. Then, use this standardised enzyme to produce your standard curve. This is the way we produce our standard curves.
Can we run the Cellazyme C tablet method (T-CCZ) at pH 6.2 instead of pH 4.5?
Modified on: Thu, 23 Sep, 2021 at 10:11 AM
Did you find it helpful? Yes No
Send feedbackSorry we couldn't be helpful. Help us improve this article with your feedback.