Acetic Acid Assay Kit (ACS Manual Format) (K-ACET)

When using the Acetic Acid Kit (K-ACET) for quantitative analysis what level of accuracy and repeatability can be expected?
The test kit is extremely accurate – at Megazyme the quality control criteria for accuracy and repeatability is to be within 2% of the expected value usin...
Fri, 25 Mar, 2016 at 12:24 PM
Is the Acetic Acid Kit (K-ACET) specific for acetate?
Acetic Acid kit (K-ACET) is specific for acetate Product Page (K-ACET)
Fri, 25 Mar, 2016 at 12:09 PM
What is the range of the Acetic Acid Kit (K-ACET)?
The assay is linear over the range of 0.3 to 20 μg of acetic acid per assay.  Using a 0.1 mL assay volume of the supplied standard at 0.1mg/mL will corre...
Fri, 25 Mar, 2016 at 12:24 PM
What is the detection limit of the Acetic Acid Kit (K-ACET)?
The detection limit is 0.14 mg/L of acetic acid, which is derived from an absorbance difference of 0.010 with the maximum sample volume of 2.0 mL. ...
Fri, 25 Mar, 2016 at 12:24 PM