Glycerol Assay Kit (K-GCROL)

When using the Glycerol Assay Kit (K-GCROL) for quantitative analysis what level of accuracy and repeatability can be expected?
The  Glycerol Assay Kit  is extremely accurate – at Megazyme the quality control criteria for accuracy and repeatability is to be within 2% of the expect...
Mon, 4 Apr, 2016 at 2:28 PM
What is the detection limit and linearity of the Glycerol Assay Kit (K-GCROL)?
The detection limit of the Glycerol Assay Kit  is 0.342 mg/L, which is derived from an absorbance difference of 0.020 with the maximum sample volume of 2...
Mon, 4 Apr, 2016 at 2:28 PM
What is the linearity of the Glycerol Assay Kit (K-GCROL)?
The Glycerol Assay Kit  is linear over the range of 0.8 to 35 μg of glycerol acid per assay.  Using a 0.1 mL assay volume of the supplied glycerol standa...
Mon, 4 Apr, 2016 at 2:29 PM
Should the absorbance (A1) values for the Glycerol Assay Kit (K-GCROL) blanks and samples be greater than 1.0?
The Glycerol Assay Kit  reaction is a negative reaction and therefore it starts with a high absorbance of ~ 1.5 and as the reaction proceeds the absorban...
Wed, 20 Apr, 2016 at 4:29 PM
Are the Glycerol Assay Kits (K-GCROL / K-GCROLGK) specific for glycerol? Will glycol contamination in my sample interfere with the assay?
The Glycerol Assay Kits (K-GCROL / K-GCROLGK) are highly specific for glycerol.  Some compounds that are known not to react or interfere with the assay incl...
Thu, 10 May, 2018 at 3:00 PM