L-Glutamic Acid Assay Kit (K-GLUT)

When using the L-Glutamic Acid Assay Kit (K-GLUT) for quantitative analysis what level of accuracy and repeatability can be expected?
The L-Glutamic Acid Assay Kit  is extremely accurate – at Megazyme the quality control criteria for accuracy and repeatability is to be within 2% of the ...
Mon, 4 Apr, 2016 at 2:19 PM
What is the detection limit of the L-Glutamic Acid Assay Kit (K-GLUT)?
The detection limit of the L-Glutamic Acid Assay Kit  is 0.214 mg/L, which is derived from an absorbance difference of 0.020 with the maximum sample volu...
Mon, 4 Apr, 2016 at 2:20 PM
What is the linearity of the L-Glutamic Acid Assay Kit (K-GLUT)?
The L-Glutamic Acid Assay Kit  is linear over the range of 0.4 to 20 μg of L-glutamic acid per assay.  Using a 0.1 mL assay volume of the supplied L-glut...
Mon, 4 Apr, 2016 at 2:21 PM
What level of cysteine in test samples will affect the results obtained from L-Glutamic Acid Assay Kit (K-GLUT)?
Samples that contain cysteine levels higher than 1 mM will not generate results within the required specification for the L-Glutamic Acid Assay Kit . ...
Mon, 4 Apr, 2016 at 2:21 PM
Absorbance values of my sample reactions continue to increase slowly after the L-Glutamic Acid Assay Kit (K-GLUT) reaction should be complete. Is there an explanation for this?
Some samples can react with the INT in the assay and cause a non-enzymatic creep reaction.  The 3rd worksheet in the MegaCalc is used to account for any c...
Mon, 4 Apr, 2016 at 2:22 PM