D-Fructose and D-Glucose Assay Kit (MegaQuant™ Format) (K-FRGLMQ)

Can oligosaccharides or polysaccharides be measured using the The D-Fructose and D-Glucose Assay Kit (MegaQuant™ Format) (K-FRGMQ)?
The kit assay will only measure free, non-covalently linked monosaccharides. Oligosaccharides or polysaccharides can be measured after hydrolysis to monos...
Fri, 1 Apr, 2016 at 10:01 AM
When using the The D-Fructose and D-Glucose Assay Kit (MegaQuant™ Format) (K-FRGLMQ) for quantitative analysis what level of accuracy and repeatability can be expected?
The test kit is extremely accurate – at Megazyme the quality control criteria for accuracy and repeatability is to be within 2% of the expected value usin...
Fri, 1 Apr, 2016 at 10:02 AM
What analyser would you recommend for use with the The D-Fructose and D-Glucose Assay Kit (MegaQuant™ Format) (K-FRGLMQ)?
The specific aim of the current assay format is to provide a simple, rugged, reliable and accurate method for the measurement of D-fructose plus D-glucose...
Fri, 1 Apr, 2016 at 10:04 AM
What is the range of the The D-Fructose and D-Glucose Assay Kit (MegaQuant™ Format) (K-FRGLMQ)?
The assay is linear over the range of 0.50 to 20 μg of D-fructose and/or D-glucose per assay.  The supplied standard of 0.5 mg/mL fructose is equivalent t...
Fri, 1 Apr, 2016 at 10:05 AM
What is the detection limit of the The D-Fructose and D-Glucose Assay Kit (MegaQuant™ Format) (K-FRGLMQ)?
The detection limit is 23.2 mg/L, which is derived from an absorbance difference of 0.02 and a sample volume of 20 μL. Product Page (K-FRGLMQ)
Fri, 1 Apr, 2016 at 10:06 AM