MegaQuant™ Wave Spectrophotometer (Chemistry Analyzer) (D-MQWAVE)

What formula is used in the installed tests on the MegaQuant Wave (D-MQWAVE) for the calculation of results?
The formula used for the MegaQaunt Wave is:  (Sample A2 - Blank A2) - (Sample A1 - Blank A1)) * Factor   As opposed to what is used in ...
Mon, 21 Nov, 2016 at 9:51 AM
How do I install the SF Capture Software that is used with the MegaQuant™ Wave Spectrophotometer (D-MQWAVE)?
Download the SF Capture software. This can be downloaded from the software link on MegaQuant™ Wave Spectrophotometer Product Page. Select (click) the down...
Thu, 10 May, 2018 at 1:17 PM