D-Sorbitol/Xylitol Assay Kit (K-SORB)

When using the D-Sorbitol/Xylitol Assay Kit (K-SORB) for quantitative analysis what level of accuracy and repeatability can be expected?
The D-Sorbitol/Xylitol Assay Kit  is extremely accurate – at Megazyme the quality control criteria for accuracy and repeatability is to be within 2% of t...
Mon, 18 Apr, 2016 at 12:10 PM
What is the detection limit of the D-Sorbitol/Xylitol Assay Kit (K-SORB)?
The detection limit of the D-Sorbitol/Xylitol assay is 0.197 mg/L of D-sorbitol and 0.165 mg/L of xylitol, which is derived from an absorbance difference ...
Mon, 18 Apr, 2016 at 12:11 PM
What is the linearity of the D-Sorbitol/Xylitol Assay Kit (K-SORB)?
The D-Sorbitol/Xylitol assay is linear over the range 1.0 to 20 µg of D-sorbitol/xylitol per assay.  Using a 0.10 mL assay volume of the supplied D-sorbit...
Mon, 18 Apr, 2016 at 12:12 PM
Why do the absorbance values of my sample reactions continue to slowly increase after the D-Sorbitol/Xylitol Assay Kit (K-SORB) reaction is complete?
Some samples can react with the INT in the assay and cause a non-enzymatic creep reaction. The 3rd worksheet in the K-SORB MegaCalc is used to accou...
Mon, 18 Apr, 2016 at 12:14 PM