The K‑ACETLQ assay kit includes a supplied standard with a concentration of 1.3 g/L.
The Sensitive Range method requires a 0.1 mL sample volume and a sample concentration between 0.033 g/L and 0.33 g/L. Since the supplied standard exceeds the upper limit of this range, it must be diluted prior to use.
To use the supplied standard with the Sensitive Range method, prepare a four‑fold dilution. For example:
Mix 0.5 mL of the supplied standard with 1.5 mL of distilled water.