Succinic Acid Assay Kit (K-SUCC)

When using the Succinic Acid Assay Kit (K-SUCC) for quantitative analysis what level of accuracy and repeatability can be expected?
The Succinic Acid Assay Kit  is extremely accurate – at Megazyme the quality control criteria for accuracy and repeatability is to be within 2% of the ex...
Mon, 18 Apr, 2016 at 3:21 PM
What is the detection limit of the Succinic Acid Assay Kit (K-SUCC)?
The detection limit of the K-SUCC Assay Kit is 0.256 mg/L which is derived from an absorbance difference of 0.010 with the maximum sample volume of 2.00 mL...
Mon, 18 Apr, 2016 at 3:22 PM
What is the linearity of the Succinic Acid Assay Kit (K-SUCC)?
The Succinic Acid assay is linear over the range 0.8 to 40 µg of succinic acid per assay. Using a 0.10 mL assay volume of the prepared succinic acid standa...
Mon, 18 Apr, 2016 at 3:23 PM
My A1 values for Succinic Acid Assay Kit (K-SUCC) blank and samples have an absorbance greater than 1.0. Is it correct for my A1 values to be this high?
The Succinic Acid Assay Kit  reaction is a negative reaction and therefore it starts with a high absorbance of ~ 1.5 and as the reaction proceeds the abs...
Mon, 18 Apr, 2016 at 3:24 PM