What is the lower limit of detection of glucose measurement using the D-Glucose Assay Kit (GOPOD Format) (K-GLUC)?

Modified on Wed, 10 Nov, 2021 at 4:41 PM

We recommend that the smallest differentiating absorbance ΔA should be at least 0.04. This is the equivalent of approximately 4 micrograms of glucose in 0.1 mL incubated with 3 mL of GOPOD Reagent.


Product Page (K-GLUC)

Was this article helpful?

That’s Great!

Thank you for your feedback

Sorry! We couldn't be helpful

Thank you for your feedback

Let us know how can we improve this article!

Select at least one of the reasons
CAPTCHA verification is required.

Feedback sent

We appreciate your effort and will try to fix the article